EZ Huber®

INSTRUCTIONS FOR USE
Use Aseptic Technique
1
Inserting
EZ Huber Safety Infusion Set
- Prepare site for injection using approved
cleaning procedures.
- Remove dust cap from female luer lock adapter
and attach to desired fluid source.
- Prime set, and expel air.
- Remove needle protector and insert needle into
skin over port. Ensure needle has reached bottom
of port. Secure set per facility protocol.
- Regulate fluid flow with fluid source.
2 Removing
EZ Huber Safety Infusion Set
- Stabilize port by placing two fingers on the
base of the EZ
Huber.
- Withdraw the needle by pulling up on the wings
of the EZ
Huber.
- The needle will slide past the safety housing.
Once the
needle passes the safety housing, the safety clip
will snap
over the needle access hole. An audible click
sound will
be heard indicating that the safety clip has
blocked the
needle access hole. The safety clip prevents the
needle
from advancing past the safety housing.
3
Disposing
EZ Huber Safety Infusion Set
• Dispose of the EZ Huber Safety
Infusion Set per institution protocol.
CT Rated for pressures to 300 psi
Single Use Only - Do not Reuse
Follow Institutional guidelines for infusion set use
Change according to CDC guidelines
Caution: Federal (USA) law restricts this device to
sale by
or on the order of a physician
STERILE: Unless package has been opened or damaged
DEHP Free
LATEX Free
pfm
Medical, Inc. Tel: 760.758.8749 2605 Temple Heights
Drive Fax: 760.758.1167 Suite A Email: customerservice@pfmmedical.com
Oceanside, CA 92056 www.pfmmedical.com
0197
pfm
Produkte für die Medizin AG Wankelstr. 60, D-50996
Köln, Germany www.pfm-ag.de
Power Injection Procedure
- Verify the implanted port is a power injection
rated port.
Access the port with the EZ Huber. Make certain
that needle tip is inserted fully within the port.
Note: Follow institutional protocol to verify
correct catheter tip position prior to power
injection.
Attach a syringe filled with sterile normal
saline.
Instruct the patient to assume the position they
will be in during the power injection procedure,
before checking for patency. If possible, the
patient should receive power injection with his or
her arm vertically above the shoulder with the palm
of the hand on the face of the gantry during
injection. This allows for uninterrupted passage of
injected contrast through the axillary and
subclavian veins at the thoracic outlet.
Aspirate for adequate blood return and
vigorously flush the port with at least 10 ml of
sterile normal saline. Warning: Failure to
ensure patency of the catheter prior to power
injection studies may result in port system failure.
Detach syringe.
Warm contrast media to body temperature.
Attach the power injection device to the EZ
Huber ensuring connection is secured. Check
indicated flow rate of safety infusion set and
confirm CT settings.
Instruct the patient to communicate immediately
any pain or change in feeling during the injection.
Inject warmed contrast, taking care not to
exceed the flow rate limits. Warning: If
local pain, swelling or signs of extravasation are
noted, the injection should be stopped immediately.
Warning: Exceeding the maximum flow rate may
result in the EZ Huber failure, implanted port
failure and/or catheter tip displacement.
Disconnect the power injection device.
Flush the EZ Huber with 10 ml of sterile normal
saline.
Perform heparin lock procedure. Note:
Some patients are hypersensitive to heparin or
suffer heparin induced thrombocytope. These patients
must not have their port primed with heparinized
saline.
After therapy completion, flush port per
institutional protocol. Close clamp while injecting
last 0.5 ml of flush solution.
Stabilize port by placing two fingers on the
base of the EZ Huber Safety Infusion Set.
Withdraw the needle by pulling up on the wings
of the EZ Huber.
The needle will slide past the safety housing.
Once the needle passes the safety housing, the
safety clip will snap over the needle access hole.
An audible click sound will be heard indicating that
the safety clip has blocked the needle access hole.
The safety clip prevents the needle from advancing
past the safety housing.
Dispose of the EZ Huber Safety Infusion Set per
institution protocol. Warning: Do not exceed
a 300 psi, or the maximum flow rate shown below, on
the power injection machine if power injecting
through the EZ Huber.
EZ Huber® is a registered
trademark of Kerr Marketing, Inc.
|
EZ Huber Gauge Size |
19G |
20G |
22G |
|
EZ Huber Clamp Color |
Brown |
Yellow |
Black |
|
Maximum Flow Setting |
5 ml/sec |
5 ml/sec |
2 ml/sec |
|
EZ Huber Gauge Size |
19G |
20G |
22G |
|
EZ Huber Clamp Color |
Brown |
Yellow |
Black |
|
Maximum Flow Setting |
5 ml/sec |
5 ml/sec |
2 ml/sec |
|
Maximum Pressure Setting |
300 psi |
300 psi |
300 psi |
|
 |
 |
 |
|
|
|
|