TUORen

TUORen
Image similar
The TUORen infusion pumps are disposable elastomeric pumps with a hard case. They are used for the application of medication. Our disposable infusion pumps have a specially designed balloon-like reservoir that is filled with the drug or fluid intended for infusion. This balloon exerts mechanical pressure there by administering the liquid based medication through a flow restrictor creating the pre-determined flow rate. The entire unit is sterile and is intended for single use only. The device is designed to deliver the nominal volume within +/- 10 % (CBI, CBI + PCA) or +/- 15 % (MR) of the nominal infusion time.

Benefits

  • Easy filling and use
  • Hard case
  • No battery or power supply required
  • Comfortable for patients


Product Details

Three types available

  • Continuous Basal Infusion (CBI type): The CBI type has a micro-flow control tube that controls the flow rate so that the medication may achieve the required therapeutic effect.
  • Patient-Controlled Analgesia (CBI+PCA type): The CBI+PCA type has an additional bolus function, so that the patient can self-administer an extra dosage according to pain situation.
  • Multirate (MR type): The MR type has an additional flow rate regulating device providing four different flow rates to choose from.

  • Accuracy: +/- 10 % (CBI, CBI+PCA) or +/- 15 % (MR)
  • Silicone reservoir allows air to dissipate
  • Extendable telescopic core
  • Particle filter
  • DEHP-free
  • Latex-free
  • Sterilized with EO valid for 5 years

Technical Data

Knowledge

The disposable infusion pumps are designed for patients requiring intravenousIntravenous (from the Latin "intra" (inner, inside) and "vena" (vein)) refers, e.g., to the administration of a medication via a vein., percutaneousPercutaneous means “performed through the skin”. (Latin: "per" - through & "cutis" - skin), subcutaneous, intraoperative sites or epidural administrationMedical term for administering or dispensing medication to patients. of medications. Medications that can be infused include analgesics, chemotherapeutics, antibiotics, antiviral drugs and infusion for re-hydration.

  • Accuracy is affected by temperature, fill volume, height and viscosity of diluent
  • Consider your local disposal guidelines

Application area

  • Analgesic therapy/PCA
  • Antibiotic treatment
  • Chemotherapy
  • Desferal therapy
  • Rehydration

Manufacturer

  • Europ. Bevollmächtigter: MedNet EC-REP GmbH, Borkstraße 10, 48163 Münster, Germany
  • Henan Tuoren Medical Device Co.Ltd., Weiyuan Industrial Zone, Menggang, Changyuan County, 453400 Henan, China

Ordering Information

Article numberBolusFlow rateInfusion timeLocking timeDelivery unit (VPE)
CBI-002-100 2 ml/h 50 h 10 pcs
CBI-003-150 3 ml/h 50 h 5 pcs
CBI-004-100 4 ml/h 24 h 10 pcs
CBI-005-060 5 ml/h 12 h 10 pcs
CBI-005-275 5 ml/h 55 h 5 pcs
CBI-010-240 10 ml/h 24 h 5 pcs
CBI-050-100 50 ml/h 2 h 10 pcs
CBI-100-100 100 ml/h 1 h 10 pcs
CBI-120-240 120 ml/h 2 h 5 pcs
CBI-200-100 200 ml/h 30 min 10 pcs
CBI-200-200 200 ml/h 1 h 5 pcs
CBI-240-240 240 ml/h 1 h 5 pcs
MR-1234-100 10 pcs
MR-2345-100 10 pcs
MR-2345-275 5 pcs
MR-2468-100 10 pcs
PCA-1020-150 2 ml 1 ml/h 6 d 8 min 5 pcs
PCA-2005-100 0.5 ml 2 ml/h 2 d 15 min 10 pcs

Downloads